Cause 1: The internal crystal temperature parameters of the laser have been changed
Solution: Contact the manufacturer to calibrate the parameters
Cause 2: Damage of the window mirror of laser
Solution: Replace the window mirror of laser
Cause 3: Unstable environment temperature
Solution: Ensure the environment temperature is between 22 to 26 degrees
Cause: Turn on the iBuild and the calibration software at the same time
Solution: Turn off the iBuild or calibration software and reconnect the laser
Cause 1: The bottom of resin level sensor is dirty
Solution: Clean the bottom of the resin level sensor
Cause 2: False angle
Solution: Adjust the angle of resin level sensor
Cause 3: Air leakage from the adsorption tube of the vacuum pump
Solution: Replace the adsorption tube
Cause 1: The volume of resin in tank is above or below standard and cannot be adjusted by the resin plunger.
Solution: Add or remove some resin. Cause 2: The resin level sensor has no signal and can’t detect the resin.
Solution: 1) Clean the bottom of the resin level sensor
2) Adjust the angle of resin level sensor
Cause: The detected laser power is not within the range of “Power Detection Maximum and Minimum RMS”.
Solution: 1) Recalibrate the laser power
2) Modify the “Target Value” or “Power Minimum RMS Value” and “Power Maximum RMS Value” of the laser power
Cause: Turn on the laser control software and calibration software synchronously
Solution: Turn off the laser control software or calibration software
Cause: Turn on the laser software and calibration software at the same time
Solution: Turn off the laser software or calibration software, then restart the iBuild.
Cause: The internal crystal temperature parameters of the laser and the environment temperature are unstable
Solution: Contact the manufacturer to calibrate the parameters and ensure that the environment temperature is between 22 to 26 degrees